Panorama Pyrometer® CellaTemp® PA 43

The Panorama Pyrometer® CellaTemp® PA 43 with an innovative optical arrangement and no moving parts comes with a rectangular measurement area. The pyrometer works according to the two-colour measuring principle; therefore, the measurement area does not have to be filled completely by the hot target object. For example, a wire may move at random within the measurement area. Yet the pyrometer provides always a secure measurement reading within milliseconds. The target marker in the through-the-lens sighting device or, when using pyrometers with a video camera, the monitor in the control room, show the exact position and size of the rectangular measurement area.

Features:

  • Range: 600 to 3000 °C
  • Focusable, interchangeable optics with rectangular measurement area
  • Precision lenses with broadband antireflection coating
  • Short wavelength spectral response
  • LED display is easy to read from a distance
  • Test current can be applied to perform diagnostics
  • Signal processing: two analogue outputs and USB / RS485 interface.
Target sighting Measuring range Lens Focus area Distance ratio / Measuring field
through-the-lens laser spot light video camera
600 – 1400 °C PZ 20.08 0.3 m – ∞ 30:1 / 150:1
650 – 1700 °C PZ 20.01 0.4 m – ∞ 45:1 / 230:1
PZ 20.03 0.2 – 0.4 m 40:1 / 215:1
PZ 20.06 1.2 m – ∞ 75:1 / 371:1
PZ 20.05 0.2 m – ∞ 10:1 / 55:1
PZ 20.08 0.3 m – ∞ 30:1 / 150:1
PA 20.06 0.6 m – ∞ 95:1 / 500:1
750 – 2400 °C PZ 20.01 0.4 m – ∞ 50:1 / 350:1
PZ 20.03 0.2 – 0.4 m 45:1 / 330:1
PZ 20.06 1.2 m – ∞ 85:1 / 580:1
PZ 20.05 0.2 m – ∞ 11:1 / 85:1
PZ 20.08 0.3 m – ∞ 34:1 / 230:1
PA 40.01 0.086-0.115 m 0.89-1.57×0.22-0.29 mm
PA 20.06 0.6 m – ∞ 105:1 / 730:1
850 – 3000 °C PZ 20.01 0.4 m – ∞ 50:1 / 350:1
PZ 20.03 0.2 – 0.4 m 45:1 / 330:1
PZ 20.06 1.2 m – ∞ 85:1 / 580:1
PZ 20.05 0.2 m – ∞ 11:1 / 85:1
PZ 20.08 0.3 m – ∞ 34:1 / 230:1
PA 40.01 0.086-0.115 m 0.89-1.57×0.22-0.29 mm
PA 20.06 0.6 m – ∞ 105:1 / 730:1
Measurement parameters
Measuring range 600 – 3000 °C in several segments
Sensor Si/Si photo diode
Spectral range 0.95 / 1.05 µm
Response time t98 Measuring range 600 – 1400 °C and 650 – 1700 °C:
≤ 10 ms (T > 750 °C)

Measuring range 750 – 2400 °C:
≤ 10 ms (T > 950 °C)

Measuring range 850 – 3000 °C:
≤ 10 ms (T > 1050 °C)

Measurement uncertainty 1.5 % of temp. reading [°C]
Repeatability 3 K
Temperature coefficient 0.05 %/K
Resolution current output 0.2 K + 0.03 % of selected range
Resolution interface 0.1 K
Resolution display 1 K
Optical variants
Lenses alle Messbereiche:
Standard lens PZ 20.01: focusing 0.4 m – ∞
Close-up lens PZ 20.03: focusing 0.2 m – 0.4 m
Weitwinkel-Objektiv PZ 20.05: Fokussierung 0,2 m – ∞
Tele-Objektiv I PZ 20.06: Fokussierung 1,2 m – ∞
Tele-Objektiv II PA 20.06: Fokussierung 0,6 m – ∞
Objektiv F50 PZ 20.08: Fokussierung 0,3 m – ∞

additionally for measuring range 750 – 2400 °C and 850 – 3000 °C:
Macro lens PA 40.01: focusing 0.086 m – 0.115 m

Electrical parameters
Analog output 0(4) – 20 mA linear, switchable, scalable
Resistance ≤ 500 Ω
Analog input 0 – 10 V
Switching output 2 x Open Collector 24 V DC, ≤ 30 mA
Switching input 2 x to 24 V
Interface USB, RS 485
Power supply 24 V DC +10 % / -20 %
Current consumption ≤ 175 mA bei 24 V DC
Camera
Video output Composite PAL, 1 Vpp, 75 Ω
Features
Sighting option Through-the-lens-sighting, laser spot light, camera
Display 4-digit LED display red, digit-height 8 mm
Status display 2 x LED
Control elements 3 buttons
General parameters
Connection M12 thread 8-pin connection
Ambient temperature 0 – 65 °C
Storage temperature -20 – +80 °C
Permissible humidity 95 % r.H. max. (non-condensing)
Dimension Ø 65 mm x 220 mm (max. 277 mm)
Housing Stainless steel V2A (1.4305)
Protection class IP 65 acc. to DIN 40050
Weight approx. 0.9 kg
Adjustable parameters
Parameters Measuring range
Measured variable and scale of analog input and output signals
Smoothing filter
Memory operation mode
Hold time of the memory
Emissivity
Transmittance
Compensation of the reflected ambient radiation
Limits and operating mode of the alarm contacts
Parameters of the ATD function (optinal)
Linearization function
Status display of the LEDs
Simulation of current and temperature
Temperature unit °C/°F
Measurement channel two-colour, single-colour 1 and single-colour 2
Threshold level of the contamination monitoring

in addition, for model with camera:
TBC function
White balance

Functions
Overtemperature signal if the internal temperature exceeds 80 °C the analog output shows a value > 20.5 mA
Service function Simulation of a measured value or a current signal
ATD-(Automatic Temperature Detection) function Automatic temperature detection at discontinuous processes
Scope of delivery
Pyrometer CellaTemp PA
Connection cable VK 02/A (5 m)
Video cable VK 02/F (5 m) for model with built-in camera
USB cable VK 11/D (1.8 m)
Software CellaView
Manual (short)
Software
CellaView
  • wires
  • rods
  • heating coils