Pyrometer CellaTemp® PA 30

CellaTemp® PA 30 is a digital pyrometer for non-contact measurement of temperatures ranging from 500 to 2500 °C. The sensor detects infrared radiation at a short wavelength which yields extremely accurate temperature readings for metals. The target can be focused either by through-the-lens sighting or with an LED spot light for target marking. The pyrometer features a rear panel display and keypad. Parameters such as emissivity, measuring range or smoothing constant can be adjusted directly at the instrument. The LED display will show parameter adjustments made using the keypad. In the normal operating mode the display indicates the current temperature reading.

Features:

  • Range: 500 to 2500 °C
  • Digital input / output
  • Focusable, interchangeable optics allow for precise distance adjustment
  • Precision lenses with broadband antireflection coating
  • LED display is easy to read from a distance
  • Test current can be applied to perform diagnostics
  • No mechanical moving parts, thus wear-free
  • Signal processing: two analogue outputs and USB / RS485 interface
Target sighting Measuring range Lens Focus area Distance ratio
through-the-lens laser spot light video camera
500 – 2500 °C PZ 20.01 0.4 m – ∞ 210 : 1
PZ 20.03 0.2 – 0.4 m 200 : 1
PZ 20.06 1.2 m – ∞ 310 : 1
PZ 20.05 0.2 m – ∞ 55 : 1
PA 20.06 0.6 m – ∞ 430 : 1
Measurement parameters
Measuring range 500 – 2500 °C
Sensor Photo diode
Spectral range 0.78 – 1.06 µm
Response time t98 ≤ 50 ms (T > 550 °C)
≤ 2 ms (T > 750 °C)
Measurement uncertainty 0.3 % of temp. reading [°C], min. 4 K
Repeatability 1 K
Temperature coefficient 0.05 %/K
Resolution current output 0.2 K + 0.03 % of selected range
Resolution interface 0.1 K
Resolution display 1 K
Optical variants
Lenses Standard lens PZ 20.01: focusing 0.4 m – ∞
Close-up lens PZ 20.03: focusing 0.2 m – 0.4 m
Weitwinkel-Objektiv PZ 20.05: Fokussierung 0,2 m – ∞
Tele-Objektiv I PZ 20.06: Fokussierung 1,2 m – ∞
Telephoto lens II PA 20.06: focusing 0.6 m – ∞
Electrical parameters
Analog output 0(4) – 20 mA linear, switchable, scalable
Resistance ≤ 500 Ω
Analog input 0 – 10 V
Switching output 2 x Open Collector 24 V DC, ≤ 30 mA
Switching input 2 x to 24 V
Interface USB, RS 485
Power supply 24 V DC +10 % / -20 %
Current consumption ≤ 175 mA bei 24 V DC
Camera
Video output Composite PAL, 1 Vpp, 75 Ω
Features
Sighting option Through-the-lens-sighting, laser spot light, camera
Display 4-digit LED display red, digit-height 8 mm
Status display 2 x LED
Control elements 3 buttons
General parameters
Connection M12 thread 8-pin connection
Ambient temperature 0 – 65 °C
Storage temperature -20 – +80 °C
Permissible humidity 95 % r.H. max. (non-condensing)
Dimension Ø 65 mm x 220 mm (max. 277 mm)
Housing Stainless steel V2A (1.4305)
Protection class IP 65 acc. to DIN 40050
Weight approx. 0.9 kg
Adjustable parameters
Parameters Measuring range
Measured variable and scale of analog input and output signals
Smoothing filter
Memory operation mode
Hold time of the memory
Emissivity
Transmittance
Compensation of the reflected ambient radiation
Limits and operating mode of the alarm contacts
Parameters of the ATD function (optinal)
Linearization function
Status display of the LEDs
Simulation of current and temperature
Temperature unit °C/°F

in addition, for model with camera:
TBC function
White balance

Functions
Overtemperature signal if the internal temperature exceeds 80 °C the analog output shows a value > 20.5 mA
Service function Simulation of a measured value or a current signal
ATD-(Automatic Temperature Detection) function Automatic temperature detection at discontinuous processes
Scope of delivery
Pyrometer CellaTemp PA
Connection cable VK 02/A (5 m)
Video cable VK 02/F (5 m) for model with built-in camera
USB cable VK 11/D (1.8 m)
Software CellaView
Manual (short)
Software
CellaView
  • metals
  • ceramics
  • high temperatures