Pyrometer CellaTemp® PA 35

CellaTemp® PA 35 is a digital pyrometer for measuring temperatures ranging between 600 and 3000 °C. The short wavelength narrow band sensor works in a spectral region where neither sight path obstructions such dust or steam nor fluctuating emissivity will interfere with the measurement result. CellaTemp® PA 35 is especially suitable for detecting the temperature of metals (Tungsten, Molybdenum, etc.) and silicon wafers. CellaTemp® PA 35 features an integrated LED display, making it easy to view the temperature reading from a distance. Use the PC USB connection to set parameters such as emissivity and smoothing time constant or to perform diagnostics. Alternatively, emissivity can be set directly at the connection cable.

Features:

  • Range: 600 to 3000 °C
  • Focusable, interchangeable optics allow for precise distance adjustment
  • Precision lenses with broadband antireflection coating
  • Selective short-wavelength narrow-band spectral response
  • Optimized to detect surface temperatures of metals and silicon wafers
  • LED display is easy to read from a distance
  • Test current can be applied to perform diagnostics
  • Signal processing: two analogue outputs and USB / RS485 interface
Target sighting Measuring range Lens Focus area Distance ratio
through-the-lens laser spot light video camera
600 – 3000 °C PZ 20.01 0.4 m – ∞ 210 : 1
PZ 20.03 0.2 – 0.4 m 200 : 1
PZ 20.06 1.2 m – ∞ 310 : 1
PZ 20.05 0.2 m – ∞ 55 : 1
PA 20.06 0.6 m – ∞ 430 : 1
Measurement parameters
Measuring range 600 – 3000 °C
Sensor Photo diode
Spectral range 0.82 – 0.93 µm
Response time t98 ≤ 50 ms (T > 650 °C)
≤ 2 ms (T > 850 °C)
Measurement uncertainty 0.3 % of temp. reading [°C], min. 4 K
Repeatability 1 K
Temperature coefficient 0.05 %/K
Resolution current output 0.2 K + 0.03 % of selected range
Resolution interface 0.1 K
Resolution display 1 K
Optical variants
Lenses Standard-Objektiv PZ 20.01: Fokussierung 0,4 m – ∞
Close-up lens PZ 20.03: focusing 0.2 m – 0.4 m
Weitwinkel-Objektiv PZ 20.05: Fokussierung 0,2 m – ∞
Tele-Objektiv I PZ 20.06: Fokussierung 1,2 m – ∞
Telephoto lens II PA 20.06: focusing 0.6 m – ∞
Electrical parameters
Analog output 0(4) – 20 mA linear, switchable, scalable
Resistance ≤ 500 Ω
Analog input 0 – 10 V
Switching output 2 x Open Collector 24 V DC, ≤ 30 mA
Switching input 2 x to 24 V
Interface USB, RS 485
Power supply 24 V DC +10 % / -20 %
Current consumption ≤ 175 mA bei 24 V DC
Camera
Video output Composite PAL, 1 Vpp, 75 Ω
Features
Sighting option Through-the-lens-sighting, laser spot light, camera
Display 4-digit LED display red, digit-height 8 mm
Status display 2 x LED
Control elements 3 buttons
General parameters
Connection M12 thread 8-pin connection
Ambient temperature 0 – 65 °C
Storage temperature -20 – +80 °C
Permissible humidity 95 % r.H. max. (non-condensing)
Dimension Ø 65 mm x 220 mm (max. 277 mm)
Housing Stainless steel V2A (1.4305)
Protection class IP 65 acc. to DIN 40050
Weight approx. 0.9 kg
Adjustable parameters
Parameters Measuring range
Measured variable and scale of analog input and output signals
Smoothing filter
Memory operation mode
Hold time of the memory
Emissivity
Transmittance
Compensation of the reflected ambient radiation
Limits and operating mode of the alarm contacts
Parameters of the ATD function (optinal)
Linearization function
Status display of the LEDs
Simulation of current and temperature
Temperature unit °C/°F

in addition, for model with camera:
TBC function
White balance

Functions
Overtemperature signal if the internal temperature exceeds 80 °C the analog output shows a value > 20.5 mA
Service function Simulation of a measured value or a current signal
ATD-(Automatic Temperature Detection) function Automatic temperature detection at discontinuous processes
Scope of delivery
Pyrometer CellaTemp PA
Connection cable VK 02/A (5 m)
Video cable VK 02/F (5 m) for model with built-in camera
USB cable VK 11/D (1.8 m)
Software CellaView
Manual (short)
Software
CellaView
  • precise measurement of metals
  • semiconductors