earthtekniks.com

CellaTemp® PA 41 is a digital pyrometer featuring fibre optics to measure temperatures ranging from 700 to 3000 °C. The sensor head is housed separately from the electronics assembly. Both the sensor head and the fibre optic cable are unsusceptible to electromagnetic radiation. The optical sensor head can withstand ambient temperatures up to 250 °C without cooling. This variant is ideal for cramped, hard-to-access locations.

Caution: The fibre optic cable must be ordered separately

Features:

  • Range: 700 to 3000 °C
  • Focusable, interchangeable optics allow for precise distance adjustment
  • Precision lenses with broadband antireflection coating
  • Short wavelength spectral response
  • LED display is easy to read from a distance
  • Test current can be applied to perform diagnostics
  • Signal processing: two analogue outputs and USB / RS485 interface.
  • Sensor head and fibre optic cable for ambient temperatures up to 250 °C
type Measuring range Lens Focus area Distance ratio
800 – 2400 °C PA 41.01 0.2 m – ∞ 190 : 1
PA 41.05 0.12 m – ∞ 100 : 1
PZ 41.18 0.033 – 0.045 m 50 : 1
900 – 3000 °C PA 41.01 0.2 m – ∞ 190 : 1
PZ 41.16 2.0 m 175 : 1
PZ 41.26 3.0 m
PA 41.05 0.12 m – ∞ 100 : 1
PZ 41.18 0.033 – 0.045 m 50 : 1
700 – 1800 °C PA 41.01 0.2 m – ∞ 100 : 1
PA 41.05 0.12 m – ∞ 50 : 1
PZ 41.18 0.033 – 0.045 m 36 : 1
1100 – 3000 °C PA 41.02 0.5 m – ∞ 340 : 1
Measurement parameters
Measuring range 700 – 3000 °C in several segments
Sensor Si/Si photo diode
Spectral range 0.95 / 1.05 µm
Response time t98 Measuring range 800 – 2400 °C:
≤ 20 ms (T > 950 °C)

Measuring range 900 – 3000 °C:
≤ 20 ms (T > 1050 °C)

Measuring range 700 – 1800 °C:
≤ 20 ms (T > 850 °C)

Measuring range 1100 – 3000 °C:
≤ 20 ms (T > 1350 °C)

Measurement uncertainty 1.5 % of temp. reading [°C]
Repeatability 3 K
Temperature coefficient 0.05 %/K
Resolution current output 0.2 K + 0.03 % of selected range
Resolution interface 0.1 K
Resolution display 1 K
Optical variants
Measuring heads Measuring range 800 – 2400 °C, 900 – 3000 °C and 700 – 1800 °C:
PA 41.01 (M30): Focusing 0.2 m – ∞
PA 41.05 (M16): Focusing 0.12 m – ∞
PZ 41.18 (M16): focusing 0.033 – 0.045 m

additionally for measuring range 800 – 2400 °C:
PZ 41.05 (M30): Fokussierung 0,3 m – ∞

Measuring range 1100 – 1300 °C:
PA 41.02 (M16): Focusing 0.5 m – ∞

Electrical parameters
Analog output 0(4) – 20 mA linear, switchable, scalable
Resistance ≤ 500 Ω
Analog input 0 – 10 V
Switching output 2 x Open Collector 24 V DC, ≤ 30 mA
Switching input 2 x to 24 V
Interface USB, RS 485
Power supply 24 V DC +10 % / -20 %
Current consumption ≤ 175 mA bei 24 V DC
Features
Sighting option Laser pilot light
Display 4-digit LED display red, digit-height 8 mm
Status display 2 x LED
Control elements 3 buttons
General parameters
Connection M12 thread 8-pin connection
Ambient temperature Electronic: 0 – 65 °C
Measuring head: 250 °C
Storage temperature -20 – +80 °C
Permissible humidity 95 % r.H. max. (non-condensing)
Dimension Electronic: Ø 65 mm x 187 mm
Measuring head: depending on the type
Housing Stainless steel V2A (1.4305)
Protection class IP 65 acc. to DIN 40050
Weight approx. 0.9 kg
Adjustable parameters
Parameters Measuring range
Measured variable and scale of analog input and output signals
Smoothing filter
Memory operation mode
Hold time of the memory
Emissivity
Transmittance
Compensation of the reflected ambient radiation
Limits and operating mode of the alarm contacts
Parameters of the ATD function (optinal)
Linearization function
Status display of the LEDs
Simulation of current and temperature
Temperature unit °C/°F
Measurement channel two-colour, single-colour 1 and single-colour 2
Threshold level of the contamination monitoring
Functions
Overtemperature signal if the internal temperature exceeds 80 °C the analog output shows a value > 20.5 mA
Service function Simulation of a measured value or a current signal
Scope of delivery
Pyrometer CellaTemp PA
Connection cable VK 02/A (5 m)
Measuring head
USB cable VK 11/D (1.8 m)
Software CellaView
Manual (short)
* The fibre optic cable must be ordered separately.
Software
CellaView
  • metals
  • cement
  • lime
  • graphite
  • glass gobs
  • crystal pulling